12.3 Recommended Safe Distance
The diode laser Medical system is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the
user of the diode laser Medical system can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the diode laser Medical system as
recommended below, according to the maximum output power of the
communications equipment.
Rated
maximum
output power
of transmitter
In Watt
Separation distance according to frequency of transmitter
In meter
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.