Page 65 of 69 USER MANUAL
12 ELECTROMAGNETIC COMPATIBILITY
Portable and mobile RF communications equipment can affect the diode laser
Medical system.
The use of ACCESSORIES, transducers and cables other than those specified, with
the exception of transducers and cables sold by the manufacturer of the diode laser
Medical system as replacement parts for internal components, may result in
increased EMISSIONS or decreased IMMUNITY of the diode laser Medical system.
The diode laser Medical system should not be used adjacent to or stacked with
other equipment and that if adjacent or stacked use is necessary, the diode laser
Medical system should be observe to verify normal operation in the configuration in
which it will be used.
Accessory part/name
Length/dimensions
Footswitch with cable
3.0m
Non-heating apparatus connecting line
2.0m
Note
Wuhan PIOON Technology Co.,Ltd. can not guarantee that accessories,
lines and transformers not delivered by Wuhan PIOON Technology Co.,Ltd.
will correspond with EMC requirements of EN 60601-1-2.
Note
The diode laser medical system is exclusively intended for use by medical
professionals. In residential areas, the diode laser medical system may
cause radio interference in certain circumstances so that it may be
necessary to undertake suitable measures such as realigning, rearranging
or screening the diode laser medical system, or filtering the connection with
the public power supply.