Installation and configuration - 15
Calibration procedure should be done by a validated sound
pressure meter.
After calibration enter correction values into to the table of
[Audiometer calibration]. Push the [OK] button to validate
changes or the [Cancel] button to abandon changes. If any
change occurred to the values the following message appears:
Do you want to save modifications?
If everything is correct, press the [OK] button. Pressing the
[Cancel] button exits without saving the changes.