EasyManuals Logo

Piston PDD-401 User Manual

Default Icon
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Installation and configuration - 15
Calibration procedure should be done by a validated sound
pressure meter.
After calibration enter correction values into to the table of
[Audiometer calibration]. Push the [OK] button to validate
changes or the [Cancel] button to abandon changes. If any
change occurred to the values the following message appears:
Do you want to save modifications?
If everything is correct, press the [OK] button. Pressing the
[Cancel] button exits without saving the changes.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Piston PDD-401 and is the answer not in the manual?

Piston PDD-401 Specifications

General IconGeneral
BrandPiston
ModelPDD-401
CategoryMeasuring Instruments
LanguageEnglish