EasyManua.ls Logo

PLUSOPTIX S16 - Page 20

PLUSOPTIX S16
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 20/22
Emissions Test
Conformity
Electromagnetic environment - Instructions
RF Emission CISPR 11
Group 1
The radiation is low and causes no interferences with
electronic devices found nearby.
RF Emission CISPR 11
Category B
The “plusoptiX S16” can be used in all areas which
are connected to the public mains supply.
Harmonic Emissions
IEC 61000-3-2
Category A
Voltage fluctuations /
Flicker IEC 61000-3-3
Complies
Immunity Test
IEC 60601 test values
Conformity values
Electromagnetic
environment -
Instructions
ESD IEC 61000-4-2
± 8 kV contact
± 2, 4, 8, 15 kV air
± 8 kV contact
± 2, 4, 8, 15 kV air
The floor should be
made of wood, ceramic
or stone. If the floor is
covered with a
synthetic material, the
relative air humidity
should not be less than
30%.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV
± 2 kV
Only plugs that are
usually available in
domestic or clinical
areas should be used.
Surge
IEC 61000-4-5
± 0.5 kV
± 1 kV
± 2 kV
± 0.5 kV
± 1 kV
± 2 kV
Only plugs that are
usually available in
domestic or clinical
areas should be used.
Voltage dips, short
interruptions on
power supply input
lines IEC 61000-4-
11
0% 0.5 cycles
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% 1 cycle
70% 25 cycles
0% 250 cycles
0% 0.5 cycles
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% 1 cycle
70% 25 cycles
0% 250 cycles
Only plugs that are
usually available in
domestic or clinical
areas should be used.
Power frequencies
and magnetic fields
IEC 61000-4-8
30 A/m, 50/60Hz
30 A/m, 50/60Hz
Magnetic fields should
not exceed the usual
areas.