Response time (time to alarm) at dose rate increase
up to 50 µR/h for 5.0±1 s produced by
137
Cs,
241
Am,
60
Co at the radiation background 20 µR/h, not more
2 s
Meets the ANSI N42.32 -2006 (6.5.1, 6.5.2)
requirements
Rate of false alarms in the mode of gamma
radiation registration at the stable radiation
background (10-20 µR/h)
mean time to false alarm > 10 h,
Meets the ANSI N42.32 -2006 (6.2.1, 6.2.2)
requirements
Response time of display indication to new
exposure rate with a relative error of indication of
±20 % at dose rate increase or decrease (from
20 µR/h to 120 µR/h; from 120 µR/h to 20 µR/h),
not more
4 s
Meets the ANSI N42.33 -2005 (6.4.1,
6.4.2) requirements
Over-range response:
the overload is indicated at the double maximum
range and remains in the maximum of the range
Meets the ANSI N42.33 -2005 (6.6.1)
requirements
Calibration by background level -аutomatic – in case the unit switching on,
change of n factors
-аutocalibration when the background
level is changed
-user initiated as needed by pressing a
button
Communication with PC via IR channel - Reading of data from storage,
- Setting of device working parameters
Power Requirements One Rechargeable Ni-MH (Nickel Metal
Hydride) Battery
1.2 V, 2100 mA/h, size AA (R6)
Battery Lifetime
up to 1.000 hours
PC Communications
IR interface
Data Collection
2.000 data points
Environmental:
- Temperature range
(LCD indication)
Humidity
- 22 to +122 °F (-30 to +50 °C)
5 to +122 °F (-15 to +50 °C)
up to 95 % at +95 °F (+35 °C)
Water Tightness
IP65
Drop Test
30” (0.7 m) onto concrete surface
Dimensions
2
13/16
x 1
1/4
” x”3
7/16
Weight, including battery 7.05 oz. (200 g.)
4.3 Battery installation
The instrument is powered by a rechargeable battery NiMH (Nickel Metal
Hydride Rechargeable battery 1.2 V min 2000 mA/h) AA size supplied with the
instrument.
The РМ1703МО-1 is packaged without a rechargeable battery installed. To
install the battery, open the cover located on the right end of the unit with a coin or
screwdriver. Once opened, insert a single rechargeable AA battery loading the positive
(+) end first. Reinstall the cover and tighten it. If the rechargeable battery is properly
installed, the LCD screen will display all segments.
6