PRK-5000 Operation Manual
- 11 -
(4) Measurement
Press the Measurement button.
Measurement will be continuously performed when the Measurement button is kept
pressed.
z The newest measured results will be displayed on the monitor.
z In case of the continuous measurement, the result of previous measurement will be
displayed.
z Just by pressing the “SIZE” button more than one second, you can select the vertex
distance. It will be displayed by SETUP when you turn on the power.
z Just by pressing the “ILLUM” button more than one second, you can change display
format of CYL value. It will be displayed by SETUP when you turn on the power.
z When measured results are displayed, pressing the “DISPLAY” button more than
one second renders each cornea measured result of R1/R2/AX Æ K1/K2/AX Æ
AR/CY/AX displayed in turn.
NOTES:
There can be error if the Outer Alignment Mark and the pupil is eccentric.
There may be some aberration to the measurement value due to the
eccentricity or inclination of lens, or deformation of cornea after surgery.
Reliability of the measurement with IOL switch turned ON may be low.