EasyManua.ls Logo

Potec PRK-5000 - Specifications

Potec PRK-5000
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PRK-5000 Operation Manual
- 40 -
8.2 Specifications
Refractometry
Vertex Distance(VD) 0.0, 12.0, 13.5, 15.0 mm
Sphere Power(SPH) -25.00 ~ +22.00 D (at the vertex distance of 12 mm)
(Increments selectable between 0.12 and 0.25 D)
Cylinder Power(CYL) 0.00 ~ ±10.00 D
(Increments selectable between 0.12 and 0.25 D)
Axis(AX) 1 ~ 180˚ (Increments: 1˚)
Cylinder Form -, +, MIX
Pupil Distance(PD) 10 ~ 88 mm
Minimum Pupil Diameter 2.0 mm
Keratometry
Radius of Curvature 5.0 ~ 10.2 mm (Increments: 0.01 mm)
Corneal Power 33.00 ~ 67.50 D (when cornea equivalent refractive index is
1.3375)
(Increments selectable from 0.05, 0.12, 0.25 D)
Corneal astigmatism 0.00 ~ -15.00 D (Increments selectable from 0.05, 0.12, 0.25 D)
Axis 1 ~ 180˚ (Increments: 1˚)
Corneal Diameter 0.8 ~ 14.0 mm (Increments: 0.1 mm)
Memory of Data 10 measured value for each right and left eye
Internal Printer Uses thermal line printer
Monitor 5.7-inch TFT LCD monitor
Environmental requirements
Operation Temperature: +10 to +40
Humidity: 30 to 85% RH
Atmospheric pressure : 70 to 106 kPa
Storage and Transportation Temperature: -10 to +55
Humidity: 10 to 95% RH
Atmospheric pressure : 50 to 106 kPa
External Interface
VGA Video Output Connected with Monitor via VGA Cable
RGB video signal plus option H and V sync
RS-232 Interface Connected with PC or PAV-6100 via RS232 Cable
RS232 Signals are r± 15[V] with Maximum 100mA
USB B Type Connected with PC via USB Cable
USB Signals are 5.00±0.25[V] with Maximum 500-900[mA]
Power supply AC100-240V, 50/60Hz
Power consumption 40 ~ 55 VA
Size
273(W)
501(D)
473(H) mm
Weight Approximately 20kg