18
30ºC (86ºF).
4. Please carefully read the label before use.
Perform Control Test
1. Insert a new test strip into the strip slot, the me-
ter will activate.
2. The last code number will appear on the
screen. Compare the code number shown on
the screen against the code number on the
test strip vial. If the two numbers match, you
may begin test, otherwise consult your local
distributor.
3. When shows up, press Set button and choose
control solution test.
4. Gently shake the control solution and apply a
drop to the aperture of strip. Make sure that
the control solution has saturated the test con-
rmation window.
5. Test result will show up in 5 seconds. The re-
sult should correspond to the range printed on
the label of strip vial used.
19
NOTE: REPEAT TEST IF THE RESULT FALLS OUTSIDE THE
CONTROL RANGE. IF YOU CONTINUE TO GET THE RESULT
FALLING OUTSIDE THE CONTROL RANGE, YOUR METER
AND STRIP MAY NOT BE WORKING PROPERLY. DO NOT
USE THE SYSTEM TO TEST YOUR BLOOD UNTIL YOU GET
A TEST RESULT FALLS WITHIN THE CONTROL RANGE.
CONSULT YOUR LOCAL DISTRIBUTOR FOR HELP.