Blood Glucose Test Strip
14
ACH
Aperture: ap-
ply the blood
from the front
strip.
Black Electrode End:
Insert this end of test
strip into meter.
Blood Reaction
Zone: It must be
filled with blood
entirely.
IMPORTANT TEST STRIP INFORMATION
1. PRECICHEK System measures the amount of
glucose in capillary whole blood. Blood can
be applied to the front of the test strip and is
15
automatically drawn to the test strip through
capillary action.
2. PRECICHEK Blood Glucose Test Strips are in-
tended for in vitro diagnostic use with capil-
lary whole blood or GL™ Control Solution. Re-
sults will not be accurate if used with plasma
or serum samples.
3. Do not use test strips beyond the expiration
date indicated on the strip vial label.
4. The discard date for test strips is 90 days af-
ter rst opening the vial. Record the discard
date on the vial when you open a new vial of
test strips.
5. Blood Glucose Test Strip can be damaged by
heat and light. Keep them sealed in the original
vial.
6.
Store the vial in a cool, dry place be-
low to 40°C/104°F and above
10°C/50°F. Do not refrigerate.