Appendix 1: Advanced Configuration
1. Commands List
Auto Adapt Light source:
First parameter: auto adapt on(1) or off(0, default)
Second parameter, only needed in case first param is 1:
desired detector level in % of saturation level (0-100), 33 is
default.
The autoadapt mode is also disabled by the LAI command!
Dispersion, only active, if SRT 0 selected (no preloaded
Index table)
0 (default): no dispersion
You should give as many Abbé numbers as there are
layers to be measured, that is (number of Peaks - 1).
The first parameter describes the nearest Layer.
rate> <effective baudrate>
<0,1>
BauD Rate for serial port
Command: “Baud rate” If 1st arg <= 8, then baudrate
index is selected, otherwise free custom baudrate is given
index of baud rate in [bit/s]
0: 9600; 1: 19200;
2: 38400; 3: 57600;
4: 115200; 5: 230400;
6: 460800; 7: 921600;
8: 1843200
2nd arg: HW Handshake on/off
Response: 2nd arg: effective baudrate (may deviate)
Continuously Refresh DarK
0 (default, no refresh)-65535 (full replacement with every
spectrum)
See also: DRK, FDK
<n>(<x>)
n: Index of the lowest measuring
rate
x: lowest frequency in Hz, floating
point
“Dark reference”
take and store dark reference
The execution of this command takes approx. 0.8sec.
The result will be valid and used for all exposure settings.
Due to the storage to nonvolatile memory, the
dark calibration will be restored on the next powerup.
See also: FDK, CRDK