40
Compatibility command 3:32Hz, 4:100Hz, 5:320Hz,
6:1000Hz, 7:2000Hz, 8: 3200Hz, 9: 4000Hz(default)
See also: SHZ
spectral d-line (587,567
nm)
Set/query Refractive Index
Only active, if SRT 0 selected (no preloaded Index table)
1.0(default)-5.0
You should give as many indices as there are layers to
be measured, that is (number of Peaks - 1). The _rst
parameter describes the nearest Layer.
See also: ABE, SRT
Set Refractive index Table
Preloaded Tables can be used instead of SRI/ABE.
SRI/ABE is active when table 0 is choosen (SRT 0)
You should give as many table indices as there are layers
to be measured, that is (number of Peaks - 1). The _rst
parameter describes the nearest Layer.
See also: SRI, ABE
2 Byte sync sequence used only in $-protocol
Saves the current setting to nonvolatile memory. The
Sensor will restart on the next power-up with this
configuration.
(default) Sending of measurement data enabled.
See also: STO
Sending of measurement data disabled. Only Command
responsesor updates are transmitted.
See also: STA
threshold for peak detection in the confocal modes (0 and
1)
0-4095, default is 50
Every exposure (or exposure burst in case of averaging)
needs a trigger event (e.g. rising edge on Trigger input).
All exposures will begin exactly at the trigger events as
long as the previous exposure is ended. If the previous
exposure is still ongoing, the trigger will be delayed. If
the device receives a trigger event, while the precedent
trigger still waits for execution, the trigger event will be
lost and the trigger lost counter will be incremented.
See also: TRG, TRW, CTN
Stops the free running measurement and waits for a
trigger event. The trigger event restarts free running