Do you have a question about the ProMinent DCM 5 Series and is the answer not in the manual?
Brand | ProMinent |
---|---|
Model | DCM 5 Series |
Category | Controller |
Language | English |
Perform an initial pH test and record the observed results before proceeding with calibration.
Press the specified buttons twice to access the pH sensor calibration menu.
Press buttons to select 'Calibrate' and 'Enter Value' for pH calibration.
Adjust the displayed pH value to match the test results and press to confirm.
Press the designated button to exit calibration and return to the main status display.
Navigate through the menu by pressing buttons to select the desired pump or sensor.
Press buttons three times to enter the editing mode for the Turn ON setpoint.
Adjust the Turn ON setpoint and press to save the new value.
Press buttons to adjust the Turn OFF setpoint and confirm the change.
Adjust the Turn OFF setpoint, which signifies feed satisfaction, and confirm.
Press the BACK button to exit setpoint adjustments and return to the main status screen.
Press once to view active alarms, showing feeder status and alarmed inputs/outputs.
Press buttons to select the desired relay (1 or 3) for mode adjustment.
Press and hold a button to display AUTO, OFF, and ON mode options.
While holding a button, use other buttons to set the output to AUTO, OFF, or ON.
Press a button, then twice, to acknowledge and clear active alarms from the system.