3 Functional description / product identification
Brief functional description
The DULCO
®
turb C is designed for the online measuring of turbidity in
untreated water, process water and treated process water in drinking water pro‐
cessing plants.
The DULCO
®
turb C product range is comprised of four device types:
Types TUC 1 and TUC 3 work with infrared light and fulfil the requirements of
the international standards ISO 7027 and DIN EN 27027.
Types TUC 2 and TUC 4 work with white light and fulfil USA standard US EPA
180.1.
Both device types can be equipped with ultrasonic cleaning (TUC 3 / TUC 4) or
without ultrasonic cleaning (TUC 1 / TUC 2). The sample cell ultrasonic cleaning
system extends the calibration and maintenance intervals in media where
coating formation is common.
DULCO
®
turb C
Part
number
IOS
7027
DIN EN
27027
US EPA
180.1
Ultrasonic cleaning
TUC 1 1037696 Infrared No
TUC 2 1037695 White
light
No
TUC 3 1037698 Infrared Yes
TUC 4 1037697 White
light
Yes
Functional description / product identification
12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCO turb C Series and is the answer not in the manual?

ProMinent DULCO turb C Series Specifications

General IconGeneral
BrandProMinent
ModelDULCO turb C Series
CategoryAccessories
LanguageEnglish

Related product manuals