The difference between the value measured by the DULCO
®
turb
C NTU-value and the value measured by the reference device is
the offset value.
For example, if the DULCO
®
turb C measures a value of 0.016
NTU for the process water, and the reference device measures a
value of 0.012 NTU for the sample, then entering an offset value of
-0.04 will cause the DULCO
®
turb C to display the value 0.012
NTU
8. At this point, the offset configuration is completed. At this stage, the device
remains in configuration mode
[CONFIG]
. Press the key
[MODE/EXIT button]
in order to return to
[AUTO]
6.4.5 Extended settings
Extended settings
A0532
Fig. 14: Extended settings
Commissioning
43