Function Options Info
Output (O/P)
n 4 - 20 mA
n 485
n off
4 - 20 mA
Value 4 mA 0 ... 1000 NTU Selection of the lower
turbidity limit value
(LOLM)✱, which corre‐
sponds to the output
value 4 mA.
Value 20 mA 0 ... 1000 NTU Selection of the upper
turbidity limit value
(UPLM)✱✱
, which corre‐
sponds to the output
value 20 mA.
✱ = a higher NTU value can be entered than for
„value 20 mA“
, in order to
reverse the sign of the output current
✱✱ = a lower NTU value can be entered than for
„value 4 mA“
, in order to
reverse the sign of the output current
485
Baud Value baud
n 1200
n 2400
n 4800
n 9600
n 19200
Selection of the baud
rate for the I/O interface
Commissioning