EasyManua.ls Logo

ProMinent DULCOTEST - 9 Technical data; 9.1 General data

ProMinent DULCOTEST
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Technical data
9.1 General data
Parameter ICT 5 ICT 5-IMA
Design For operation in the flow For immersion
Measuring principle inductive
Measuring range 0 ... 2000 mS/cm
Precision within the measuring range
0 ... 1 mS/cm ≤ 1%
0 ... 10 mS/cm ≤ 0.5%
0 ... 50 mS/cm ≤ 0.5%
0 ... 100 mS/cm ≤ 0.5%
0 ... 1000 mS/cm ≤ 1%
0 ... 2000 mS/cm ≤ 1%
Cell constant k = 6.25 1/cm
Temperature sensor Pt1000, Class A
t
90
Temperature
a
≤ 6 seconds
Permissible ambient temperature -10 ... +60 °C
Permissible storage temperature -20 ... +75 °C
Degree of protection
b
IP68
Permitted medium temperature
when operating in pipework made
of:
PVC: -10 ... +60 °C
PP: -10 ... +80 °C
-10 ... +60 °C
Maximum process pressure with PP design
at 20 °C 10 bar 0 bar
at 60 °C 6 bar 0 bar
at 80 °C 0 bar
Technical data
30