7.3.3 LIGHT CURVE (LC) PROTOCOL
The protocol called Light Curve (LC) was designed to acquire parameters for construction of Light Response Curve
relating the rate of photosynthesis to photon flux density. The method is based on successive measurements of the
sample exposed to a stepwise increase of light intensity. The effective quantum yields of photosynthesis are determined
under various light intensities of continuous illumination. Measurement is based on pulse modulated fluorometry (PAM).
Several LC protocols are predefined in FP. These differ in number and duration of individual light phases and light
intensities.
# of phases
Phase duration
Light intensities [µmol.m
-2
.s
-1
]
LC1
6
60s
10; 20; 50; 100; 300; 500
LC2
5
30s
100; 200; 300; 500; 1000
LC3
7
60s
10; 20; 50; 100; 300; 500; 1000
Tab. 3 LC Protocols.
The protocol includes following measured and calculated parameters:
Abbreviation
Explanation
F
0
minimum fluorescence in dark-adapted state
F
m
maximum fluorescence in dark-adapted state
F
m
_Ln
maximum fluorescence in light adaptation state
F
t
_Ln
instantaneous fluorescence during light adaptation
QYmax*
maximum quantum yield of PSII in dark-adapted state - Fv/Fm
QY_Ln
**
instantaneous PSII quantum yield induced in light
n represents a sequential number of light phase
*Calculated as (F
m
F
0
) / F
m
** Calculated as (F
m
_Lx F
t
_Lx) / F
m
_Lx

Other manuals for PSI FluorPen FP 110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PSI FluorPen FP 110 and is the answer not in the manual?

PSI FluorPen FP 110 Specifications

General IconGeneral
BrandPSI
ModelFluorPen FP 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals