PT200MI instruction manual 24
8.2.3 SETTING THE COMPARATOR SET-POINTS
All data relates to the value displayed in accordance with the setting of function F19.
Note the position of the decimal point and the division prior to entering setting mode as the value
entered during setting should be in the same format as that set for display during function setup. i.e.
if the scale has been set to show “100.00” at maximum capacity and the division is 5 then a set-point
of 76.05 would be entered as 7605. The decimal point does not show as it is used to indicate the digit
being set.
8.2.4 SETTING SET-POINTS
Enter the comparator setting mode: Press and hold MODE , then press the ZERO key, 2 seconds later
the PT200MI displays “SET”, a further 2 seconds later the gross annunciator begins flashing and the
PT200MI displays “000000”
Step 1: HIGH HIGH set-point setting (HH)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the gross annunciator flashes.
Accept the data by pressing the MODE key.
Step 2: HIGH set-point setting (HI)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the net annunciator flashes.
Accept the data by pressing the MODE key.
Step 3: LOW set-point setting (LO)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the motion annunciator flashes.
Accept the data by pressing the MODE key.
Step 4: LOW LOW set-point setting (LL)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the zero annunciator flashes.
Accept the data by pressing the MODE key.
Step 5: HIGH HIGH hysteresis set-point setting (HH-S)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the gross annunciator flashes.
Accept the data by pressing the MODE key.
Step 6: HIGH hysteresis set-point setting (HI-S)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the net annunciator flashes.
Accept the data by pressing the MODE key.
Step 7: LOW hysteresis set-point setting (LO-S)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the motion annunciator flashes.
Accept the data by pressing the MODE key.
Step 8: LOW LOW hysteresis set point setting (LL-S)
Select the digit to change by pressing the ZERO key and set the value by pressing the TARE key.
During this step the zero annunciator flashes.
Accept the data by pressing the MODE key.
Step 9: Exit the comparison setting
Press MODE key to return to normal weighing mode.
An error will occur if the following setting condition is not satisfied: HH>HI>LO>LL,
if an error occurs press the MODE key to redo from steps 1 to step 4.
Note: If one of the set-point value is zero then the value will not be used for comparison. For example:
HH = 00, then “HH” is inactive.