PT200MI instruction manual 9
5.1 DATA SETTING
Entering the data setting mode:
• Press and hold the MODE key, then press the G/N key, 2 seconds later, “FUNC” displays.
• A further 2 seconds later “F0 0” is displayed.
• Press ZERO to move to the next setting.
• To exit the setting mode press ZERO until “F0 0” is displayed and then press MODE to save
the function settings.
Setting data:
Data selection is performed with the MODE key.
• Continue pressing MODE until the number indicating the desired selection appears.
• The ZERO key accepts the setting and moves on to the next function.
• If no change is required press the ZERO key to move to the next function.
• To save the data press ZERO until "F0 0" is reached and then press MODE.
5.1.1 "F0 0" EXIT THE FUNCTION SETTING MODE
Press MODE key to accept and save the data for all the functions altered, and enter the weighing
mode.
Press ZERO key to skip to the next setting.
5.1.2 "F1 0" ZERO TRACK TIME
Press TARE key to select, ZERO to accept.
0 = No track 1 = 1 sec
5.1.3 "F2 0" ZERO TRACK BAND
Press TARE key to select, ZERO to accept.
0 = 1 step width 1 = 2 step width 2 = 4 step width
MULTIPLIER F4 0.1 (F4=2) 1 (F4=1) 10 (F4=0)
DIVISION F5 1 2 5 1 2 5 1 2 5
(F5=0) (F5=1) (F5=2) (F5=0) (F5=1) (F5=2) (F5=0) (F5=1) (F5=2)
Increment or step 0.1 0.2 0.5 1 2 5 10 20 50
500 (F6=0) 50.0 50.0 50.0 500 500 500 ERR ERR ERR
1000 (F6=1) 100.0 100.0 100.0 1000 1000 1000 1000 ERR ERR
1200 (F6=2) 120.0 120.0 120.0 1200 1200 1200 1200 ERR ERR
1500 (F6=3) 150.0 150.0 150.0 1500 1500 1500 1500 ERR ERR
2000 (F6=4) 200.0 200.0 200.0 2000 2000 2000 2000 2000 ERR
2500 (F6=5) 250.0 250.0 250.0 2500 2500 2500 2500 2500 ERR
3000 (F6=6) 300.0 300.0 300.0 3000 3000 3000 3000 3000 ERR
MAXIMUM 4000 (F6=7) 400.0 400.0 400.0 4000 4000 4000 4000 4000 ERR
CAPACITY 5000 (F6=8) 500.0 500.0 500.0 5000 5000 5000 5000 5000 5000
F6 6000 (F6=9) 600.0 600.0 600.0 6000 6000 6000 6000 6000 6000
8000 (F6=10) 800.0 800.0 800.0 8000 8000 8000 8000 8000 8000
10000 (F6=11) 1000.0 1000.0 1000.0 10000 10000 10000 10000 10000 10000
12000 (F6=12) ERR 1200.0 1200.0 ERR 12000 12000 12000 12000 12000
15000 (F6=13) ERR 1500.0 1500.0 ERR 15000 15000 15000 15000 15000
20000 (F6=14) ERR 2000.0 2000.0 ERR 20000 20000 20000 20000 20000
25000 (F6=15) ERR ERR 2500.0 ERR ERR 25000 25000 25000 25000
30000 (F6=16) ERR ERR 3000.0 ERR ERR 30000 30000 30000 30000
40000 (F6=17) ERR ERR 4000.0 ERR ERR 40000 40000 40000 40000
50000 (F6=18) ERR ERR 5000.0 ERR ERR 50000 50000 50000 50000
60000 (F6=19) ERR ERR ERR ERR ERR ERR 60000 60000 60000
80000 (F6=20) ERR ERR ERR ERR ERR ERR 80000 80000 80000
100000 (F6=21) ERR ERR ERR ERR ERR ERR 100000 100000 100000