PT200MI instruction manual 10
5.1.4 "F3 0" MOTION DETECTION
Press TARE key to select, ZERO to accept.
0 = 1 step width/sec 1 = 3 step width/sec 2 = 5 step width/sec 3 = 10 step width/sec
"F4 0" MULTIPLIER OR DECIMAL POINT
Press TARE key to select, ZERO to accept.
0 = 10 1 = 1 2 = 0.1 3 = 0.01 4 = 0.001 5 = 0.0001
5.1.6 "F5 0" DIVISION
Press TARE key to select, ZERO to accept.
0 = 1 1 = 2 2 = 5
5.1.7 "F6 0" MAXIMUM CAPACITY
Press TARE key to select, ZERO to accept.
0 = 500 1 = 1000 2 = 1200 3 = 1500 4 = 2000 5 = 2500
6 = 3000 7 = 4000 8 = 5000 9 = 6000 10
= 8000 11 = 10000
12 = 12000 13 = 15000 14 = 20000 15 = 25000 16 = 30000 17 = 40000
18 = 50000 19 = 60000 20 = 80000 21 = 100000
If the maximum capacity does not satisfy the following condition, it will show
“ERROR 1” for 2 seconds then return to the F4 (decimal point) setting mode.
Maximum capacity >= 500
Maximum capacity <= 100,000
Note: “ERROR 1” is also output to RS232/RS485 if the option is fitted.
5.1.8 "F7 0" BAUD RATE
Press TARE key to select, ZERO to accept.
0 = 2400 baud 1 = 4800 baud 2 = 9600 baud 3 = 19200 baud
5.1.9 "F8 0" RS232/RS485 OUTPUT MODE
Press TARE key to select, ZERO to accept.
0 = Continuous mode (weight continuously output)
1 = Command mode (weight output in response to READ instruction over RS232/RS485)
5.1.10 "F9 0" UNIT
Press TARE key to select, ZERO to accept.
0 = kg 1 = ton
5.1.11 "F10 0" ZERO RETURN RANGE
Press TARE key to select, ZERO to accept.
0 = 1% 1 = 2% 2 = 3% 3 = 4% 4 = 5% 5 = 6% 6 = 7% 7 = 8% 8 = 9% 9 = 10%
Note: In the normal weighing mode, the PT200MI can be zeroed with a reading within the range 1% to
10% of maximum capacity by pressing ZERO key.
5.1.12 "F11 0" DIGITAL FILTER
Press TARE key to select, ZERO to accept.
0 = 0 1 = 2 2 = 4 3 = 8 4 = 16 5 = 32 6 = 64 7 = 128 8 = 256 9 = 512
5.1.13 "F12 0" DISPLAY UPDATE RATE
Press TARE key to select, ZERO to accept.
0 = 1 time/sec 1 = 4 times/sec 2 = 8 times/sec 3 = 16 times/sec 4 = 20 times/sec