PT200MI instruction manual 3
2. SPECIFICATIONS
2.1 GENERAL
1. Mains supply : 110VAC, 220VAC±10%, 50/60Hz
2. Power consumption : 9 VA
3. Operating temperature : -5ºC to 50ºC(23ºF to 122ºF)
4. Relative Humidity : 90%(non-condensing)
5. Weight : Approx 0.45 kg
2.2 DIGITAL SECTION
1. Weight display : 6 digits LED display
2. Display height : 13mm
3. Annunciators : Gross, Net, Zero, Motion and unit(kg/t)
4. Negative sign indication : “-” on the left most digit
5. Over-range indication : Displays “O.L” or “-O.L”
6. Maximum capacity : 500 to 100,000
7. Step width : 1, 2, 5, 10, 20 or 50
8. Decimal point : Displays to four different decimal places
2.3 ANALOGUE SECTION
1. Load cell type : All strain gauge load cells
2. Load cell supply : 10VDC±5%, 150mA
3. Input sensitivity : 0.5μV/D to 200μV/D, max. 20mV input
4. Input resistance : More than 100MΩ at 500VDC between
each terminal
5. Zero point adjustment : 0.05mV to 15mV
6. Span stability : ±8ppm/ K of F.S
7. Zero stability : ±(0.4μV±0.006% initial zero offset
voltage)/K
8. Non-linearity : Within 0.005% of F.S
9. Conversion method : Delta-sigma
10.Conversion rate : Up to 200 times/sec
11.Internal resolution : 16,000,000
12.Maximum display resolution : 10,000 divisions
13.Comparison cycle : Approx 200 times/sec
14.Dielectric strength
-Between input terminal(common/earth/each opto-coupler
output/analogue output/BCD output), for 1 min. : 500VDC
-Between power supply terminal and input
terminal(common/earth/each opto-coupler
output/analogue output/BCD output), for 1 min. : 1500VAC
2.4 RS232/RS485 SERIAL OUTPUT (2 OPTIONS)
1. Standard : EIA-RS232/RS485 output
2. Stream format : 7 Data bits, 1 stop bit, even parity
a) Baud rate : 2400, 4800, 9600, 19200
b) Data type : ASCII code
c) Delimiter : CR/LF
3. Communication mode
a) Command mode:
-After receiving the command word from the RS232/RS485 port, the indicator will carry
out the appropriate action. Those commands valid only for RS232/RS485 port are.
Command: READ<CR><LF> : request measured data
TARE<CR><LF> : request TARE weight setting
KEY<CR><LF> : key protection ON or OFF
ZERO<CR><LF> : request ZERO value
Example: Read command (READ<CR><LF>) is “52H, 45H, 41H, 44H, 0DH, 0AH”
in ASCII code, numbers in hexadecimal.
b) Continuous mode:
-The data will be transmitted constantly without any input command to the RS232/RS485 port.