Page 24
P084 Jump CutOff (Software Revisions from 3.6.0 to 6.3.7)
This parameter determines the distance, from the face of the transducer, that the
unit will process the First valid echo, providing that a Larger echo is not seen
beyond this point.
Values: Min. 0.300
Max. 99.999
P084 Mid Loss (from software revisions 6.8.?)
This Parameter now determines the height in milli-volts of the point where the
near loss becomes the far loss i.e. the loss change point (P083)
Loss Limit Values
Default = 40.0m
Default = 15.00 (150mv)
P41 Ringdown Loss
P81 Near Loss
P83 Mid Loss
(Vertical Movement)
P82 Far Loss