- 96 -
FIG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, filter symbols
Example:
Command:
FIG<CR><LF>
- give current
filter
Response:
FIG_2_OK<CR><LF>
- currently set
average filter
Set Value release
Format: ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
command carried out
ARS_E <CR><LF>
error in-course of command execution, no parameter or
incorrect format
ARS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining value release options
n → 1 fast
2 fast+reliable
3 reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release settings
are assigned to the working mode
An example:
Command:
ARS_2<CR><LF>
set value release parameter to
fast+reliable option
Response:
ARS_OK<CR><LF>
fast+reliable option set
Give current value release
Command overview:
On receipt of a command, the terminal sends information on currently value release.
Format:
ARG <CR><LF>
Response options:
ARG_x_OK<CR><LF>
- command carried out, returns set value release
ARG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, value release symbol
Example:
Command:
ARG<CR><LF>
- give current value release
Response:
ARG_2_OK<CR><LF>
- currently set value release parameter to
fast+reliable option
Set last digit
Format: LDS_n <CR><LF>
Response options:
LDS_OK <CR><LF>
command carried out
LDS_E <CR><LF>
error in-course of command execution, no parameter or incorrect format
LDS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining last digit settings
n → 1 always
2 never
3 when stable
CAUTION!
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit settings
are assigned to the working mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS PLUS Series and is the answer not in the manual?

RADWAG AS PLUS Series Specifications

General IconGeneral
DisplayLCD with backlight
Power Supply100-240 V AC 50/60 Hz
CalibrationInternal / External
InterfaceRS232, USB-A, USB-B
Operating Temperature+10° - +40°C
Pan Sizeø 85 mm
Repeatability0.1 mg
Linearity±0.2 mg
Stabilization time2-3 seconds