- 92 -
Set ambient conditions state
Format: EV_n CR LF
Response options:
EV_OK CR LF
command carried out
EV_E CR LF
error in-course of command execution, no parameter or incorrect format
EV_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining ambient conditions state
n 0 unstable ambient conditions
1 stable ambient conditions
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
EV_1 CR LF
set value ‘stable’ for ambient conditions option
Response:
EV_OK CR LF
ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps it to
value <UNSTABLE>.
Set filter
Format: FIS_n CR LF
Response options:
FIS_OK CR LF
command carried out
FIS_E CR LF
error in-course of command execution, no parameter or incorrect format
FIS_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining filter number
n → 1 very fast
2 fast
3 average
4 slow
5 very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.
An example:
Command:
FIS_3 CR
LF set average filter
Response:
FIS_OK CR
LF average filter set
Set Value release
Format: ARS_n CR LF
Response options:
ARS_OK CR LF
command carried out
ARS_E CR LF
error in-course of command execution, no parameter or incorrect format
ARS_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining value release options
n → 1 fast
2 fast+reliable
3 reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release
settings are assigned to the working mode
An example:
Command:
ARS_2 CR LF
set value release parameter to fast+reliable option
Response:
ARS_OK CR LF
fast+reliable option set

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g

Related product manuals