EasyManuals Logo

RADWAG AS R2 User Manual

RADWAG AS R2
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
- 93 -
Set last digit
Format: LDS_n CR LF
Response options:
LDS_OK CR LF
command carried out
LDS_E CR LF
error in-course of command execution, no parameter or incorrect format
LDS_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining last digit settings
n → 1 always
2 never
3 when stable
CAUTION!
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit settings
are assigned to the working mode.
An example:
Command:
LDS_1 CR LF
set last digit option to value ‘always’
Response:
LDS_OK CR LF
‘always’ value set
Cooperation with PUE 7.1, PUE 10 terminal
Format: NT CR LF
Response options:
ES CR LF
- command not understood (format error)
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2
3
4
5
6
7
8
9-18
19
20-22
23
24-32
33
34-36
37
38
39
40
N
T
space
stability
marker
zero marker
range
marker
digit marker
space
mass
space
mass unit
space
tare
space
tare unit
space
Hidden
digits
quantity
CR
LF
NT
- command
Stability marker
- [space] if stable, [?] if unstable
Zero marker
- [space] for any value but zero, [Z] for zero value
Range marker
- marker informing on range within which mass value is comprised:
[space] I range balance, [2] II balance, [3] III balance
Digit marker
[zero] no digit, [1] one digit marker, [2] two digits marker, [3] three
digits marker, [4] four digits marker, [5] five digits marker
Mass
10 characters for net mass given in adjustment unit (with dot and a
floating-point “-“ sign )
Mass unit
3 characters with left justification
Tare
9 characters with dot - right justification (when floating-point tare gets
turned off automatically then zero value is sent)
Tare unit
3 characters with left justification
Hidden digits quantity
Number specifying quantity of hidden digits: [space] when no digits
are hidden, [1] when 1 digit is hidden
An example:
NT CR LF
- command sent from a computer
N T _ ? _ _ 0 _ _ _ _ _ - 5 . 1 1 3 _ g _ _ _ _ _ _ _ 0 . 0 0 0 _ g _ _ _0 CR LF command carried
out, response: mass value and remaining data
Where: _ - space

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g