49
18.3. Introducing reference mass to balance memory
Mass of single piece sample or multiple can be introduced to assortment base
in following way:
a) determine reference mass (point 7.2 a, point 7.2 b)
b) enter assortment base F or F2 (PC keyboard)
c) choose assortment from assortment base (set indicator at given
assortment)
d) press INSERT key
Reference mass will be saved for given assortment under position “Pcs. mass"
18.4. Counting pieces with checkweighing
Fig. 34. Counting pieces with result control
1 - control indicator
LO- to small quantity of pieces (Min. Pcs. in assortment base)
OK.- value in range ( Min. Pcs. Max. Pcs. in assortment base)
HI- to big quantity (value bigger than Max. Pcs. parameter in assortment base)
To activate Result control set Aut. Printout parameter in Operating settings
on YES.
NOTICE:
For each assortment there must be Min pcs. and Max pcs. entered in
assortment base.