79
Standard length is 18 characters, but if the parameter digit marking has been
enabled, the length increases to 20 characters, because additional brackets
are printed e.g.:
383 kg
38[3] kg
3[83] kg
It can be set in :
Settings/Factory/digit marking
Frame format:
[space] [sign] [mass] [space] [unit] [CR LF]
sign – [space] for positive values or [-] for negative values
mass – 9 characters justified to the right
unit – 3 characters justified to the left
This type of printouts can be enabled in parameters:
Settings/RS-232/Aut. print/when stab.
Setting the Min. mass is possible in:
Settings/RS-232/Min. mass
Notice:
The result is printed in current unit.
26.4.5. Survey of automatic printouts
No Name Frame
Quantity of
characters
0 No
[Command] [space] [stability] [space] [sign]
[mass] [space] [unit] [CR LF]
20-22
1
Continu
ous
[SI] [space] [stability] [space] [sign] [masa]
[space] [unit] [CR LF]
21
2
Wit
interv.
[SI] [space] [stability] [space] [sign] [masa]
[space] [unit] [CR LF]
21
3
When
stab.
[space] [sign] [space] [mass] [space] [unit]
[CR LF]
18

Table of Contents