4.4.1.2 Rules for ripple test definition
The analyzer places few restrictions on the definition of ripple limit ranges.
The following rules ensure a maximum of flexibility:
Ranges do not have to be sorted in ascending or descending order (e.g. the "Start
Stimulus" value of range no. n does not have to be smaller than the "Start Stimu-
lus" value of range no. n+1).
Overlapping ranges are allowed. The limit check in the overlapping area is related
to the tighter limit (the pass test involves a logical AND operation).
Gaps between ranges are allowed and equivalent to switching off an intermediate
ripple limit range.
Ripple limit ranges can be partially or entirely outside the sweep range, however,
the limits are only checked at the measurement points.