Concepts and features
R&S
®
ZNA
260User Manual 1178.6462.02 ─ 20
●
Controlled timing is limited to setups with a single channel.
●
Controlled timing is not compatible with:
– ALC
– spectrum measurements
– two-dimensional compression point measurements
– phase coherent and true differential signal generation
– external devices control via USB or LAN (power meters, generators)
– RFFE usage
4.7.11  Noise figure measurement
Option R&S ZNA-K30
Software option R&S ZNA-K30 allows you to measure the noise figure of a 2-port DUT,
operating in its linear range.
Currently, this option is only available for R&S ZNA26 and R&S ZNA43.
Noise figure (or noise factor, if the linear representation is chosen) is an important fig-
ure of merit for system designers and eventually determines the overall system perfor-
mance of a receiver.
Figure 4-27: Exemplary spectrum for a single tone excitation (left = DUT input; right = DUT output)
The general definition of the noise factor F
D
 is the ratio of the signal to noise ratio
(SNR) at the input, to the SNR at the output of a DUT. In other words, it describes the
degradation of the SNR from the input to the output of a DUT.
Figure 4-28: Schematic showing the signal and noise powers relevant for noise figure calculations
Optional extensions and accessories