EasyManua.ls Logo

R&S ZNA50 - Calculate:fmodel:sfd<Ph_Pt>:Preset; Calculate:fmodel:sfd<Ph_Pt>:Run:run

R&S ZNA50
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
958User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:SFD<Ph_pt>:PRESet
Restores the default settings of the SFD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Reset to Default" on page 675
CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]
<Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the test fixture is connected.
This has to be defined before measuring DUT + Fixture (using CALCulate:FMODel:
SFD<Ph_pt>:DUT:MEASure).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> 1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Manual operation: See "Active" on page 678
CALCulate:FMODel:SFD<Ph_pt>:RUN:RUN
Runs the SFD tool.
Before executing this command, make sure that:
The test coupon measurement (using CALCulate:FMODel:SFD<Ph_pt>:
COUPon:MEASure) finished successfully, or the test coupon data were success-
fully loaded from file (using CALCulate:FMODel:SFD<Ph_pt>:COUPon:
MEASure:FILename
The measurement of DUT + test fixture (using CALCulate:FMODel:
SFD<Ph_pt>:DUT:MEASure) finished successfully.
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\ZNA\Embedding. If result files with the same name already exist, they are
overwritten.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Event
Manual operation: See "Run <Fixture Modeling Tool>" on page 679