GUI reference
R&S
®
ZNA
675User Manual 1178.6462.02 ─ 20
Suppress Remeasure Warnings
If checked, the Measure Coupon and Measure DUT + Test Fixture do not raise warn-
ings, if the measurement is repeated with different ports.
Remote command:
n.a.
Reset to Default
Restores the default settings of the selected fixture modeling tool. This comprises:
the states and settings in the Measure Coupon, Measure DUT + Test Fixture and
Save Fixture Model sections (common for all tools)
the advanced settings of the selected tool (tool specific)
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:PRESet
CALCulate:FMODel:SFD<Ph_pt>:PRESet
CALCulate:FMODel:EZD<Ph_pt>:PRESet
Apply
This button is only active, if the selected deembedding tool has been successfully run
(see Chapter 5.14.5.4, "Save Fixture Model", on page 679).
Applies the calculated deembedding files <...>_left_DUT.sNp and
<...>_right_DUT.sNp) to the ports marked "Active" in the Save Fixture Model sec-
tion.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<LogPt>
Cancel
Closes the dialog without applying any deembedding files.
5.14.5.2 Measure Coupon
The "Measure Coupon" section allows you to measure one or more test coupons for
the related fixture.
Offset Embed softtool