GUI reference
R&S
®
ZNA
676User Manual 1178.6462.02 ─ 20
Coupon Type
Selects the coupon type to be measured. The following coupon types are supported:
ISD SFD EZD
"Sym 2x Thru"
"1x Open"
"1x Short"
"1x Open, 1x Short"
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:COUPon:TYPE
CALCulate:FMODel:SFD<Ph_pt>:COUPon:TYPE
Measure / Measure Open / Measure Short
Starts the coupon measurement at the Active ports.
The display area shows all the S parameter measurements being made. The resulting
Touchstone file is written to C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Embed-
ding.
For Coupon Type "1x Open, 1x Short" (ISD only) the measurement proceeds in two
steps: one for the "1x Open" and one for the "1x Short" coupon. The label of the "Mea-
sure" button changes accordingly.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure:OPEN
CALCulate:FMODel:ISD<Ph_pt>:COUPon:MEASure:SHORt
CALCulate:FMODel:SFD<Ph_pt>:COUPon:MEASure
CALCulate:FMODel:EZD<Ph_pt>:COUPon:MEASure on page 939