EasyManua.ls Logo

R&S ZNA50 - Page 959

R&S ZNA50
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
959User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:SFD<Ph_pt>:RUN[:STATe] <Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the deembedding files shall be assigned after the tool has run.
Suffix:
<Ph_pt>
.
Physical port number for single-ended deembedding, logical port
number for balanced deembedding.
Parameters:
<Boolean> 1 (ON) if a deembedding file shall be assinged to port <Pt>, 0
(OFF) otherwise
Manual operation: See "Apply" on page 680
CALCulate:FMODel:SFD<Ph_pt>:TOTaldiffcfg <SFDPortConfig>
Tells the SFD tool about the port ordering of the test fixture.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<SFDPortConfig> ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
Manual operation: See "Total Port Ordering" on page 686
More...
CALCulate:FMODel:DIRectory <String>
Defines the common working directory of the ISD, SFD and EZD fixture modeling tools.
Parameters:
<String> Directory path, either absolute or relative to the current directory
(MMEMory:CDIRectory).
Manual operation: See "Output Settings" on page 674
CALCulate:FMODel:DIRectory:DEFault
Resets the common working directory of the ISD, SFD, and EZD tool to its default
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Embedding.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Output Settings" on page 674
SCPI command reference