EasyManua.ls Logo

R&S ZNA50 - Page 686

R&S ZNA50
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
686User Manual 1178.6462.02 ─ 20
Total Port Ordering
Tells the SFD tool about the port ordering of the test fixture
"Ports 1&3 on Left" (default): odd ports are on the left and even ports are on the
right
"Ports 1&2 on Left" ports 1 to N are on the left and ports N+1 to 2·N are on the
right
Remote command:
CALCulate:FMODel:SFD<Ph_pt>:TOTaldiffcfg
Adjust Impedance Mismatch
If checked, the SFD tool performs automatic impedance adjustments.
Remote command:
CALCulate:FMODel:SFD<Ph_pt>:AUTO on page 955
5.14.5.8 EZD Advanced Settings
Allows you to specify advanced parameters for the EZD tool.
Port Order
Tells the EZD tool about the port ordering of the test coupon
"Odd on left" (default): odd ports are on the left and even ports are on the right
"1 to N on left" ports 1 to N are on the left and ports N+1 to 2·N are on the right
Remote command:
CALCulate:FMODel:EZD<Ph_pt>:PORT:ORDer
Generate Side 1/Generate Side2
If checked (default), the tool generates the deembedding file for side 1 / side 2.
Offset Embed softtool