GUI reference
R&S
®
ZNA
345User Manual 1178.6462.02 ─ 20
IP<order><side><DUT port>
Intercept Point at DUT output
●
<order> defines the order of the intercept point. Possible values are 3, 5, 7, or 9.
●
<side> defines the position of the intercept point relative to the lower and upper
tones.
– The Lower intercept points are measured at frequencies below the lower tone.
– The Upper intercept points are measured at frequencies above the upper tone.
– The Minor intercept point denotes the lower or upper intercept point, whichever
is smaller. It reveals the worst-case performance of the DUT.
●
<DUT port> defines the reference port for the intercept point calculation. Possible
values are O for the DUT output port and I for the DUT input port.
Both values differ by the attenuation of the lower tone signal upon transmission
through the DUT; see "Intermodulation quantities" on page 241.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
Main Tone tab
Allows you to select the upper/lower tone at DUT input/output as measured quantities.
Access: Trace – [Meas] > "Intermodulation" > "Main Tone"
Meas softtool