GUI reference
R&S
®
ZNA
662User Manual 1178.6462.02 ─ 20
5.14.2.1 Controls on the Offset tab
Overview
This button is available on all "Offset Embed" softtool tabs. It opens the Overview panel
in the Offset Embed dock widget.
Port
Physical test port of the analyzer. You can define independent offset parameters for all
ports.
Remote command:
The <PhyPt> numeric suffix in the [SENSe<Ch>:]CORRection:... commands
identifies the physical port.
Delay / Electrical Length / Mech. Length
Defines the length offset at the selected port as a delay, an electrical length, or a
mechanical length. The three quantities are related by:
and overwrite each other. See also Chapter 4.6.1.1, "Definition of offset parameters",
on page 201.
Note: The entered parameters must correspond the actual (one-way) length of the
transmission line. To account for the propagation in both directions, the phase shift of a
reflection parameter due to a given length offset is twice the phase shift of a transmis-
sion parameter. For a numeric example, see Chapter 4.6.1.6, "Application and effect of
offset parameters", on page 206.