EasyManua.ls Logo

R&S ZNA50 - Page 90

R&S ZNA50
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Concepts and features
R&S
®
ZNA
90User Manual 1178.6462.02 ─ 20
Parallel Measurements with Frequency Offset
When performing parallel measurements, it is possible to specify a minimum frequency
offset between the port groups. This is particularly useful for situations where "Cross-
talk" between different DUTs would otherwise make the simultaneous measurement
impossible (e.g. in wafer prober applications).
Example:
Consider a linear frequency sweep with 101 points from 100 MHz to 200 MHz resulting
in a frequency step size of 1 MHz. When two port groups are defined and a minimum
frequency offset of 1 MHz is specified, a parallel measurement with frequency offset is
performed internally as follows:
In the first measurement step, no measurements for port group 1 will be performed;
measurements for port group 2 will be performed at 100 MHz.
In the second measurement step measurements for port group 1 will be performed
at 100 MHz; measurements for port group 2 will be performed at 101 MHz.
In the third measurement step measurements for port group 1 will be performed at
101 MHz; measurements for port group 2 will be performed at 102 MHz.
In the 101st measurement step measurements for port group 1 will be performed at
199 MHz; measurements for port group 2 will be performed at 200 MHz.
Finally, in the 102nd measurement step measurements for port group 1 will be per-
formed at 200 MHz; no measurements for port group 2 will be performed.
Parallel measurement with frequency offset is transparent to the user: All port groups
will be measured in the requested frequency range. The results are available in the
same form as if they were obtained in separate measurements without frequency off-
set.
Please note that in parallel measurement with frequency offset the firmware uses a
modified IF as compared to measurements not using this mode. Because this modified
IF requires a special calibration, it is essential to perform the Calibration with the same
Frequency Offset settings as for the actual measurement; otherwise the calibration is
turned off (see "Cal Off label" on page 160).
4.1.4.3 Stimulus and sweep types
The function of the Stimulus keys [Start], [Stop], [Center] and [Span] depends on the
sweep type.
Table 4-1: Function of Stimulus keys
Sweep type [Start](unit) [Stop] (unit) [Center] (unit) [Span] (unit)
"Lin Freq" "Start Frequency" (Hz) "Stop Frequency" (Hz) "Center Frequency"
(Hz)
"Span Frequency" (Hz)
"Log Freq" "Start Frequency" (Hz) "Stop Frequency" (Hz)
"Segmented"
Basic concepts

Table of Contents

Other manuals for R&S ZNA50

Related product manuals