Command reference
R&S
®
ZNA
964User Manual 1178.6462.02 ─ 20
Manual operation: See "Auto Linearity Deviation" on page 429
CALCulate<Chn>:LDEViation:CONStant <Constant>
Defines the constant value for the linearity deviation calculation.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Constant> Constant value
Range: The range depends on the trace format. For the
default trace: -200 dB to + 200 dB.
Increment: 10E-2 dB
*RST: 0 dB
Example: See CALCulate<Chn>:LDEViation:AUTO on page 963
Manual operation: See "Constant" on page 430
CALCulate<Chn>:LDEViation:ELENgth <ElLength>
Defines the electrical length for the linearity deviation calculation. This command is
available if the active trace format is Phase or Unwrapped Phase, or if a mixer delay is
measured.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<ElLength> Electrical length
Range: 0 m to + 1000 m.
Increment: 10E-4 m
*RST: 0 m
Default unit: m
Example: See CALCulate<Chn>:LDEViation:AUTO on page 963
Manual operation: See "El. Length" on page 430
CALCulate<Chn>:LDEViation:MODE <Mode>
Applies /discards the correction factors or re-calculates them for each trace (Tracking).
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Mode> OFF | ON | TRACking
Apply the correction using the latest correction factors.
ON - Apply the correction using the latest correction factors.
OFF - Discard the correction.