EasyManua.ls Logo

R&S ZNA50 - Page 981

R&S ZNA50
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
981User Manual 1178.6462.02 ─ 20
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
Manual operation: See "Limit Check" on page 432
CALCulate<Chn>:LIMit:STATe:AREA <HorizontalPos>, <VerticalPos>
Moves the limit check pass/fail message for the active trace <Chn> to one of nine pre-
defined positions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example: See CALCulate<Chn>:LIMit:FAIL?
Manual operation: See "Limit Check" on page 432
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe] <Boolean>
Switches the TTL pass/fail signals on or off. The signals are applied to the USER
PORT as long as the active trace <Chn> is within limits, including the ripple limits.
See Chapter 12.3.1.1, "User Port", on page 1644.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Pt> 1 - TTL out pass 1 (pin 13 of User Port connector)
2 - TTL out pass 2 (pin 14 of User Port connector)
Parameters:
<Boolean> ON | OFF - TTL output signal on or off.
*RST: OFF
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; TTL2 ON
Switch the limit check on and activate the TTL out pass 2 signal.
Manual operation: See "TTL1 Pass / TTL2 Pass" on page 434
SCPI command reference

Table of Contents

Other manuals for R&S ZNA50

Related product manuals