45
WR14 / WR50-12/13 Winding Resistance Meter Instruction Manual
Version 3.03
MODE
Single
Performs 1 single measurement.
Stop condition:
Stops the measurement as soon as the result is
within the selected conditions.
(Example Shown: within ±0.10% during 10 Seconds
of time)
Continuous
Performs continuous measurement until the
measurement is stopped by the user.
Quality Assistant:
If the result is within the selected parameters, a
green light in the main screen will illuminate and
indicate this condition.
If the result is out of the selected parameters, a red
light in the main screen will illuminate and indicate
this condition.
If the selected conditions are almost met, a yellow
light in the main screen will illuminate.