Version 3.03
50
The correction formula:
𝑅
𝑠
= 𝑅
π‘š
𝑇
𝑠
+ 𝑇
π‘˜
𝑇
π‘š
+ 𝑇
π‘˜
Rs Resistance corrected
Rm Measured resistance
Ts Desired reference temperature
Tm The temperature at which resistance was/is measured
Tk 234.5 for copper
225.0 for aluminum
(aluminum may be as high as 240)
Alternative: Correction with β€œalpha25” to calculate Tk with the formula:
𝑇
π‘˜
=
1
π‘Žπ‘™π‘β„Žπ‘Ž25
βˆ’ 25
When temperature correction is activated, the Main Test Screen appears as below:
The β€œT” in the measuring channel box shows that
the displayed resistance value is corrected. In this
example it is corrected to 25 Β°C using Tk of Copper
and referring to Temperature probe 1.