EasyManua.ls Logo

Raytech WR14 - Page 54

Raytech WR14
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
WR14 / WR50-12/13 Winding Resistance Meter Instruction Manual
Version 3.03
Details
After selecting results, press Details and view the
screen with details of the selected measurements.
If no measurements were selected, the details of all
stored measurements are displayed.
Return to the results overview by selecting
"Header".
Caption: This is the caption (name) of the measurement channel.
Rmeas: This is the measurement Result without a temperature correction.
Rcorr: This is the temperature corrected measurement result.
T1-T3: The temperature of the probes.
Text: Manual input of temperature value (measured with an external temperature meter).
Scroll to the right to view additional details.
Itest: The measurement current.
Ch: The channel which measured the result.
Time: Time between the start of the measurement and memory storage.

Table of Contents

Related product manuals