WR14 / WR50-12/13 Winding Resistance Meter Instruction Manual
Version 3.03
Pin 5 and Pin 6 is part of the extended safety circuit. When these contacts are open, the instrument acts
as if the emergency stop has been pressed. Pin 5 and 6 must be closed with an external switch or
shorting plug.
Install the shorting plug, shown here,
to use the instrument without the
extended safety circuit.
Pin 5 and Pin 6 are shorted together in
the plug.
Extended Safety Circuit Type Connector with 6 pins:
Panelmount Connector on the instrument:
Lemo: ERA.1E.304.CCL
RAN: 40802
Cablemount Connector (Extern):
Lemo: FFA.1E.306.CLAC60
RAN: 40803