Installation
MI 15
Figure4:ProperSensorPlacement
5.2Wiring
5.2.1SensorHeadCable
Themanufacturer preinstall’s the sensor head cable betweensensor
headandelectronicbox.Itmaybeshortenedbutnotlengthened.
Shortening the cable length by 1 m (3 ft.) causes a
temperatureerrorof0.1K/m!
Do not bend the sensor head cable tighter
than 25 mm / 1 in.
(MID/MIC)and15mm/0.6in.(MIH)respectively!
Target greater than spot size
Target equal to spot size
Target smaller than spot size
best
good incorrect
Sensor