Maintenance
MI 61
9.2Fail‐SafeOperation
TheFail‐Safesys temisdesignedtoalerttheoperatorandprovidea
safe outputincaseof any system failure. The sensoris designed to
shutdowntheprocessintheeventofaset‐uperror,systemerror,ora
failureinthesensorelectronics.
The Fail‐Safe circuit should
never be relied on
exclusively to protect critical processes. Other safety
devicesshouldalsobeusedtosupplementthisfunction!
When an error or failure does occur, the display indicates the
possible failure area,andtheoutputcircuits automaticallyadjustto
theirpresetlevels,seethefollowingtables.
ErrorCodes
fortheOutputs
Symptom mV 0 to 20 mA 4 to 20 mA TC-K TC-J
Temperature over
range
5 V 21 mA 21 mA - -
Temperature under
range
0 V 0 mA 2.5 mA - -
Defect of the internal
head ambient
temperature probe
5 V 21 mA 21 mA > 1200°C