ProgrammingGuide
72 MI
11.4.3PostProcessi ng
Thefollowingparameterscanbesettodeterminethepostprocessing
mode,seesection
6.4PostProcessingonpage38.
P=5 peakhold,holdtime:5s
F=12.5 valleyhold,holdtime:12.5s
G=10 averaging,averagetime(90%):10s
XY=3 advancedpeakhold,hysteresis:3K
XY=2 advancedvalleyhold,hysteresis:2K
AdvancedPeak/ValleyHoldwithAveraging:
C=250 threshold:250°C
AA=15 averagingtime(90%):
15s
11.5DynamicData
Alltemperaturerelatedinformationiscalculated128timesasecond.
Torequestthedynamicdata,followingcommandsareavailable:
?T targettemperature
?I internaltemperatureofthesensinghead
?XJ internaltemperatureofelectronicshousing
?Q energyvalueoftheinfraredtemperature
?XT triggersetpoint(active/inactive)fortheFTC3
input
To check for resets (e.g. power shut down ) use the command XI.
Notice,afteraresettheunitisnewinitialized.
?XI asksfortheresetstatus
!XI0 noresetoccurred
!XI1 aresetoccurred,newinitializationoftheunit
XI=0 setstheresetstatusbackto0