ProgrammingGuide
MI 71
Thereareeightentriespossibleforemissivitysetting(1)andarelated
setpoint(threshold)(2).Tobeabletowriteorreadthesevalues,use
thefollowingcommands:
EP=2 setpointerfortableentry,e.g.toline2(3)
RV=0.600 settheemissivityvalueforline2to0.600(4)
SV=220.0 setthesetpoint(threshold)forline2to220.0(5)
Figure46:TableforEmissivityandSetPoints
Toactivatetheseemissivitysettings,youneedtohavethe3external
inputs(FTC)connected.Accordingtothedigitalcombinationonthe
FTCwires,oneofthetableentrieswillbeactivated,seesection
5.4.2
EmissivitySettingviaDigitalSelectiononpage25.
1
2
3
4
5