Installation
16 MI
5.2.2CableforPowerSupplyandOutputs
Youneedtoconnectthepowersupply(12to26VDC)andthesignal
outputwires.Useonlycablewithoutsidediameterfrom4to6mm
(0.16to0.24in),AWG24.
The cable must include shielded wires. It should not be
used
asastrainrelief!
1. Cutabout40mm(1.5in)ofthecablesheath(7)fromtheend
ofthecable.Caution:Donotcutintotheshield!
2. Cut the shield (5) so about 5 mm (0.2 in) remains exposed
from under the cable sheath (7). Separate the
shield and
spread the strands out. Shorten the inside insulation until
youcanseparatethewires(6).
3. Strip3mm(0.15in)ofinsulationfromthewires(6).
Figure5:CablePreparation
4. Open the electronicbox by removing the four Phillips head
screws and pulling off the lid. Unscrew the cap (1), and
removetheplasticcompressionfitting(2),therubber washer
(3),whichisinsidethefitting,andthetwometalwashers(4).