EasyManua.ls Logo

RayTek MI Miniature Infrared Sensor - Lock Mode; Mode Setting for the Digital Input FTC3; Changing the Sensing Head Calibration Data74; Ambient Background Temperature Compensation

RayTek MI Miniature Infrared Sensor
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProgrammingGuide
74 MI
XF factorydefaultvalueswillbeset
11.6.5LockMode
The access to the unit is possible via serial interface (software) and
viathedirectuserinput(mode buttons,LCDdisplay).Itispossible
tolockthebuttons.Thisallowsaccesstheunitonlyviasoftware.
J=L directuserinputvia
modebuttonsdenied
Alternatively the unit can be unlocked by pressing the <Mode/Up>
buttonssimultaneouslyforthreeseconds.
11.6.6ModeSettingfortheDigitalInputFTC3
ThedigitalinputFTC3(seesection
5.4.4TriggerandHoldFunction
onpage28)canbeusedasfollows:
XN=T FTC3astrigger
XN=H FTC3withholdfunction
11.6.7ChangingtheSensingHeadCalibrationData
If it is necessary to exchange the sensing head, you must set the
calibrationdataforthenewsensinghead:
XZ=01234567FFFFFFFF accordingtothe
headcalibrationdata
ForMID/MICmodelsthelasteightnumbersarenotusedandmust
besetto„F“likeshownintheexampleabove.ForMIHmodelsand
specials(likeMTBorG5)allsixteennumbersareused.
11.6.8AmbientBackgroundTemperatureCompensation
In case of compensating the ambient
background temperature, the
followingmodesareavailable:
AC=0 nocompensation
AC=1 compensation with a constant temperature value set
withcommandA.