ASCII Programming
126 Rev. G Nov/2015 MI3
Description
Char
Format
P
B
S
N
Legal Values
Factory default
Head /
Box
Sensor initialization
XI
1 after reset
0 if XI=0
1
B
Box Temperature
XJ
float
(C/F)
B
FTC3 trigger/hold
XN
nXN
T = trigger, H = hold
T
H
analog Output 1 mode
XO1O
(XO)
1
XO1O=v
(XO=v)
0 0...20 mA
2
4 4...20 mA
5 TCJ
3
6 TCK
7 TCR
8 TCS
9 0...5 V
10 0...10 V
99 disable (tristate)
XO1O=9 for
communication
box (metal)
XO1O=99 for
communication
box 6TE, analog
B
analog Output 2 mode
XO2O
XO2O=v
0 0...20 mA
4 4...20 mA
9 0...5 V
10 0...10 V
99 disable (tristate)
XO2O=4 for
communication
box (metal)
XO1O=99 for
communication
box 6TE, analog
B
analog Output 3 mode
4
XO3O
XO3O=v
0 0...20 mA
4 4...20 mA
9 0...5 V
10 0...10 V
99 disable (tristate)
XO3O=99 for
communication
box 6TE, analog
B
analog Output 4 mode
5
XO4O
XO4O=v
0 0...20 mA
4 4...20 mA
9 0...5 V
10 0...10 V
99 disable (tristate)
XO4O=99 for
communication
box 6TE, analog
B
Box Firmware Revision
XR
?XR
e.g. 1.01
Set in FW
B
Setpoint relay function
XS
nXS
float
°C/°F
500°C
H
Trigger
XT
?XT
0 = inactive, 1 = active
0
B
Unit identification
XU
?XU
e.g. !XUMI3COMM
set at production
Serial number
XV
?XV
e.g. !XV98123
set at production
B
Advanced hold
hysterese
XY
nnnn
H
Timer counter (if V=B)
Indicates the Burst
speed
6
Z
$=..Z..
-
0 9999 ms (repeat)
--
B
1
for MI3COMM only
2
Mode 0 and 4 for MI3MCOMMA only
3
Mode 5 to 8 for MI3COMM only
4
for MI3MCOMMA only
5
for MI3MCOMMA only
6
Burst mode only